# Diaconn G8 Insulin Pump ## Insulin Pump Bluetooth Pairing - Click on the hamburger menu in the top left corner. ![image](../images/DiaconnG8/DiaconnG8_01.jpg) - Click on Config Builder. ![image](../images/DiaconnG8/DiaconnG8_02.jpg) - After selecting the Diaconn G8 Pump click on the Settings icon (cog wheel). ![image](../images/DiaconnG8/DiaconnG8_03.jpg) - Choose Selected pump. ![image](../images/DiaconnG8/DiaconnG8_04.jpg) - Select your insulin pump’s model number once it appears in the list. ![image](../images/DiaconnG8/DiaconnG8_05.jpg) - There are two options to check your model number: 1. The last 5 digits of the SN number on the back of the pump. 2. Click on O button > Information > BLE > Last 5 digits. ![image](../images/DiaconnG8/DiaconnG8_06.jpg) - Once you select your pump, a window appears asking for a pin code. Enter the pin number displayed on your pump to complete the connection. ![image](../images/DiaconnG8/DiaconnG8_07.jpg) ## Pump status check and log synchronization - Once your pump is connected, click on the Bluetooth symbol to check the status and to synchronize logs. ![image](../images/DiaconnG8/DiaconnG8_08.jpg) ## Bluetooth Troubleshooting **What to do in the case of an unstable Bluetooth connection with the pump.** ### Method 1 ) Check the pump again after AAPS application is completed. - Click on the 3 dots button on the top right. ![image](../images/DiaconnG8/DiaconnG8_09.jpg) - Click on Exit. ![image](../images/DiaconnG8/DiaconnG8_10.jpg) ### Method 2) If the first method doesn’t work, disconnect Bluetooth and then reconnect. - Press and hold the Bluetooth button at the top for about 3 seconds. ![image](../images/DiaconnG8/DiaconnG8_11.jpg) - Click on the Setting button on the paired Diaconn G8 Insulin pump. ![image](../images/DiaconnG8/DiaconnG8_12.jpg) - Unpair. ![image](../images/DiaconnG8/DiaconnG8_13.jpg) - Repeat the Bluetooth pairing process for the pump (see above). ## Further Information ### Diaconn G8 Insulin pump option setting - Config manager > pump > Diaconn G8 > Settings - DIACONN G8 at the top> 3 dots button on the top right > Diaconn G8 Preferences ![Diaconn G8 pump options](../images/DiaconnG8/DiaconnG8_14.jpg) - If the **Log reservoir change** option is activated, the relevant details are automatically uploaded to the careportal when an “Insulin Change” event occurs. - If the **Log needle change** option is activated, the relevant details are automatically uploaded to the careportal when a “Site Change” event occurs. - If the **Log tube change** option is activated, the relevant details are automatically uploaded to the careportal when a “Tube Change” event occurs. - If the **Log battery change** option is activated, the relevant details are automatically uploaded to the careportal when a “Battery Change” event occurs, and the PUMP BATTERY CHANGE button in the ACTION tab is deactivated. (Note: To change the battery, please stop all in-progress injection functions before proceeding.) ![Diaconn G8 actions menu](../images/DiaconnG8/DiaconnG8_15.jpg) ### Extended Bolus function - If you use extended bolus it will disable closed loop. - See [this page](Extended-Carbs-why-extended-boluses-won-t-work-in-a-closed-loop-environment) for details why extended bolus does not work in a closed loop environment.